Advanced Patient Monitor (10/12/15 inch )
-The monitor adopts Ergonomic design and a compact structure to meet the needs of various parameter detections. It is suitable for adults, children, and newborns.
-Standard parameters: SpO2, 3/5 leads ECG, PR, HR, RESP, NIBP, 2*TEMP.
-Optional parameters: EtCO2, 12 leads ECG, 2*IBP.
-10/12/15 inch color LCD screen with optional touch screen and rotator knob.
-Capable to visualize 6 or more waveform. Support color coding for different waveform.
-28 ST segment analysis with ST trend with continuous display of ST value on screen.
-Multiple screen layout to view big font size in numeric and waveform.
-Trending facilities for up to 360 hours of both graphical & numerical along with event review facility for all parameters and full waveform disclosure facility.
-Different visual alarm with color coding and audible alarms for various parameters with three levels of
volume adjustment.
-Support HL7 protocol to access the central monitoring system, low-power design, extending battery life Configure, fanless design to ensure a quieter and cleaner environment Arrhythmia detection and analysis in various leads.
-Standard application: 5-leads ECG, NIBP, SpO2, RESP, TEMP, PR/HR
-Optional application: Touch screen, IBP, EtCO2 (mainstream/sidestream), 12-lead ECG, recorder, CMS
-Large font, dynamic short trend display interface, calculation of hemodynamics and cardiac output.
Size: 10 inches Color TFT LCD 12 inches Color TFT LCD 15 inches Color TFT LCD
Resolution: 1024*768 pixels 1366*768 pixels
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
ECG Range: 0~100%
ECG Resolution: 2.36uV/LSB
HR Range: 15~300bpm(adult), 15~250bpm(pediatric/neonate) HR Accuracy: ±1 bpm or ±1%, whichever is greater.
RR Range: 0~120bpm
RR Accuracy: 15~120rpm: ±2rpm or ±2% <15rpm: Undefined
Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s, less than ±5% error
Display sensitivity: 2.5 mm/mV (X0.25), 5 mm/mV (X0.5), 10 mm/mV (X1), 20 mm/mV (X2), 40 mm/mV (X4), Auto, less than ±5% error
SPO2 Measurement Range: 0~100%
SPO2 Accuracy: 70 to 100%: ±2% (adult/pediatric, without motion)
70 to 100%: ±3% (neonate, without motion)
0 to 69%: Not specified.
PI Range: 0~20%
PR Range: 25-250bpm
PR Accuracy: ±2bpm or ±2% (whichever is greater)
Technique: Thermal resistance
Measurement Range: 0~50.0℃ Accuracy: ±0.2°C
Resolution: 0.1°C Refresh Rate: 1s
Alarm limit range: Temp High: (low limit +0.1℃) ~ 50.0℃
Temp Low: 0℃ ~ (high limit -0.1℃)
Measure method: Automatic oscillation method Pressure Range: 0~300mmHg (0~40kPa)
Pressure Accuracy: ±2mmHg or ±1% of reading (whichever is greater) Resolution: 1mmHg or 0.1kPa
SYS Range: Adult: 40-270mmHg
Pediatric: 40-200mmHg Neonate: 40-130mmHg
DIA Range: Adult: 10-210mmHg
Pediatric: 10-162mmHg Neonate: 10-90mmHg
Mean Range: Adult: 20-230mmHg
Pediatric: 20-170mmHg Neonate: 20-100mmHg
Accuracy: mean deviation<±5mmHg, standard deviation<8mmHg Work pattern: manual, automatic, continuous measurement
Automatic mode measurement interval: 1min/ 2min/ 3min/ 4min/ 5min/ 10min/ 15min/ 30min/ 60min/ 90min/ 2h/ 3h
Continuous mode measurement time: 5 min Maximum single measurement time: 180s
Technique: Trans-thoracic impedance Lead: Options are lead I and II
Respiration excitation waveform: <300µA RMS, 62.8kHz (±0%) Respiration impedance range: 0.3~5Ω
Baseline impedance range: 200~2500Ω (using an ECG cable with 1kΩ resistance) Bandwidth: 0.2~2Hz (-3 dB)
Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s
Measurement range: 25bpm~250bpm Resolution: 1bpm
Alarm limit: PR High: (PR Low limit +1bpm) ~ 300bpm PR Low: 15bpm ~ (PR High limit -1bpm)
Measurement range: 15~300bpm (adult)
15~350bpm (pediatric/neonate)
Resolution: 1bpm
Accuracy: ±1 bpm or ±1%, whichever is greater Sensitivity: 0.2mV
Trends: A minimum of 120 hours trend data with the resolution no less than 1 minute Events: 200 events
NIBP measurements: 200 events
Interpretation of resting 12-lead ECG results: 20 sets
Number of channels: 2
Pressure Range: -50~350mmHg (-6.7~40.0kPa)
Accuracy: ±2% or ±1 mmHg, whichever is greater (excluding sensor error) Resolution: 1 mmHg (0.1kPa)
Refresh Rate: 1s Sensitivity: 5μV/V/mmHg
Zero adjustment range: ±200 mmHg Impedance range: 300~3000Ω
Volume displacement: <0.04mm³/100mmHg
Measurement mode: Sidestream, Mainstream Technique: Infrared absorption
Zero RR delay: 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s CO2 Range: 0~20.0 vol%
CO2 Accuracy: 0~12 vol%: ± (0.2 vol%+2% of readings) 12~20 vol%: ± (0.2 vol%+6% of readings)
AwRR Measurement Range: 0~150rpm Resolution: 1 rpm
AwRR Accuracy: Mainstream: 0~150rpm, ±1rpm
Sidestream: 0~69rpm, ±1rpm; 70~150rpm, Undefined Alarm limit: EtCO2 High: (EtCO2 Low limit +1mmHg) ~ 150mmHg
EtCO2 Low: 0 ~ (EtCO2 High limit -1mmHg) FiCO2 High: 0 ~ 150mmHg
AwRR High: (AwRR Low limit +1 rpm) ~ 120rpm AwRR Low: 0 ~ (AwRR High limit -1 rpm)
The multi-parameter monitor is composed of host, battery, display screen, parameter module, parameter module plug-in box and accessories (pulse saturation probe, body temperature probe, ECG lead, blood pressure cuff, sampling tube). This monitor supports hot plugging for Plug-In Module, where they can be plugged without turning off the monitor.
Standard Para: lead ECG, RESP, SpO2, 2-TEMP, NIBP, PR/HR
Optional Para: 2-IBP, EtCO2, touch screen, recorder, 12-lead ECG, cardiac output, anesthesia depth EEG, CMS
-Suitable for adults, children, and newborns.
-12.1/15.6 inches color LCD screen with optional touch screen and rotator knob.
-Upgradable module is interchangeable (plug and play type).
-Capable to visualize 6 or more waveform.
-Support color coding for different waveform.
-28 ST segment analysis with ST trend with continuous display of ST value on screen.
-Multiple screen layout to view big font size in numeric and waveform.
-Trending facilities for up to 360 hours of both graphical & numerical along with event review facility for
all parameters and full waveform disclosure facility.
-Different visual alarm with color coding and audible alarms for various parameters with three levels of volume adjustment.
-Support HL7 protocol to access the central workstation
-Energy-saving, low-power design, extending battery life
-Configure noise reduction fan to ensure a quieter and cleaner environment Arrhythmia detection and analysis in various leads.
-5200mAh built-in battery backup of 4 hours
-Dimension: 30cm*15cm*30cm
Size:
12.1 inches Color TFT LCD 15.6 inches Color TFT LCD Resolution:
1024*768 pixels 1366*768 pixels
ECG Range: 0.15~5.5mV ECG Resolution: 2.36uV/LSB
HR Range: 15~300bpm (adult) 15~350bpm (child/neonate) HR Accuracy: ±1bpm or ±1% (whichever is greater)
RR Range: 0~120bpm
RR Accuracy: 15~120rpm: ±2rpm or ±2% <15rpm: Undefined
Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s, less than ±5% error
Display sensitivity: 2.5 mm/mV (X0.25), 5 mm/mV (X0.5), 10 mm/mV (X1), 20 mm/mV (X2), 40 mm/mV (X4), Auto, less than ±5% error
SPO2 Range: 0~100%
SPO2 Accuracy: 70~100%, ±2% <70%, Undefined PI Range: 0~20%
PVI Range: 0.001%
PR Range: 25-250bpm
PR Accuracy: ±2bpm or ±2% (whichever is greater)
Technique: Thermal resistance Measurement Range: 0~50.0℃ Accuracy: ±0.2°C
Resolution: 0.1°C
Refresh Rate: 1s
Alarm limit range: Temp High: (low limit +0.1℃) ~ 50.0℃
Temp Low: 0℃ ~ (high limit -0.1℃)
Technique: Oscillometry
Mode of operation: Manual, Auto, STAT, Sequence Pressure Range: 0~300mmHg
Pressure Accuracy: ±2mmHg or ±1% of reading (whichever is greater) Resolution: 1mmHg
SYS Range:
Adult: 40-270mmHg Pediatric: 40-200mmHg Neonate: 40-135mmHg
DIA Range:
Adult: 10-210mmHg Pediatric: 10-150mmHg Neonate: 10-100mmHg
Mean Range:
Adult: 20-230mmHg Pediatric: 20-165mmHg Neonate: 20-110mmHg
Accuracy: the mean deviation<±5mmHg, the standard deviation<8mmHg
Technique: Trans-thoracic impedance Lead: Options are lead I and II
Respiration excitation waveform: <300µA RMS, 62.8kHz (±0%) Respiration impedance range: 0.3~5Ω
Baseline impedance range: 200~2500Ω (using an ECG cable with 1kΩ resistance) Bandwidth: 0.2~2Hz (-3 dB)
Sweep speed: 6.25 mm/s, 12.5mm/s, 25mm/s, 50mm/s
Measurement range: 25bpm~250bpm Resolution: 1bpm
Accuracy: ±3bpm
Alarm limit: PR High: (PR Low limit +1bpm) ~ 300bpm PR Low: 15bpm ~ (PR High limit -1bpm)
Measurement range: 15~300bpm (adult)
15~350bpm (pediatric/neonate)
Resolution: 1bpm
Accuracy: ±1 bpm or ±1%, whichever is greater Sensitivity: 0.2mV
Trends:
60 hours trend data with the resolution no less than 1 minute. 4 hours trend data with the resolution no less than 5 second. Events: 100 events
NIBP measurements: 40 events
Interpretation of resting 12-lead ECG results: 25 sets
Pressure Range: -50~350mmHg (-6.7kPa~40.0kPa)
Accuracy: ±2% or ±1 mmHg, whichever is greater (excluding sensor error) Resolution: 1 mmHg (0.1kPa)
Sensitivity: 5μV/V/mmHg
Zero adjustment range: ±200 mmHg Impedance range: 300~3000Ω
Volume displacement: <0.04mm³/100mmHg
Range: 0.20~20.00L/Min Accuracy: ±5%
ETCO2 (Optional) CO2 Range: 0~20.0vol% CO2 Accuracy:
0~12 vol%: ± (0.2 vol%+2% of reading) 12~20 vol%: ± (0.2 vol%+6% of reading) AwRR Range: 0~150rpm
Resolution: 1 rpm AwRR Accuracy:
Mainstream: 0~150rpm, ±1rpm
Sidestream: 0~69rpm, ±1rpm; 70~150rpm, Undefined
1. 8-inch, two kinds of configuration for choice: 3 Para, or 6 Para 3 Para: NIBP, SPO2, PR
6 Para: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR
2. 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
3. Efficient resistance to interference of defibrillator and electrosurgical cautery;
4. SPO2 can testing for 0.1% Weak;
5. RA-LL impedance Respiration;
6. Trend Coexist Display;
7. OxyCRG DyNamic View Display;
8. Bed to Bed view Display;
9. Networking capacity and nurse calling system;
10. Options of Central Monitoring System;
11. Built-in rechargeable battery 2200mAh;
12. 8" high resolution color TFT LCD display;
13. Large volume of tabular and graphic trends information storage and easy to recall;
14. Anti-ESU, anti- defibrillator;
15. Capture dynamic waveforms.
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V, Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL) Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Range
Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which great Resolution: 1 bpm
Sensitivity: > 200 (uV P-P) Differential Input Impedance: > 5 MΩ CMRR
-Monitor: > 105 dB
-Operation: > 105 dB
-Diagnosis: > 85 dB
Electrode offset potential: ±300mV Leakage Current: < 10 uA
Baseline Recovery: < 3 S after Defi. ECG Signal Range: ±8 m V (Vp-p) Bandwidth
-Surgery: 1 ~ 15 Hz
-Monitor: 0.5 ~ 35 Hz
-Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available Review: Available
Method: Impedance between R-F (RA-LL) Differential Input Impedance: >2.5 MΩ Measuring Impedance Range: 0.3~5.0Ω Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth: 0.3 ~ 2.5 Hz Resp. Measuring Range:
-Adult: 0 ~ 120 rpm
-Neo/Ped: 0 ~ 150 rpm Resp. Alarm Range:
-High alarm range limit: 8~120 rpm
-Low alarm range limit: 6~120 rpm Resolution: 1 rpm
Accuracy: ±2 rpm
Apnea Alarm: 10 ~ 40 S
Method: Oscillometric Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min Pulse Rate Range: 40 ~ 240 bpm Alarm Type: SYS, DIA, MEAN Measuring and alarm range
Adult Mode
-SYS: 40 ~ 270 mmHg
-DIA: 10 ~ 215 mmHg
-MEAN: 20 ~ 235 mmHg
Pediatric Mode
-SYS: 40 ~ 200 mmHg
-DIA: 10 ~ 150 mmHg
-MEAN: 20 ~ 165 mmHg
Neonatal Mode
-SYS: 40 ~ 135 mmHg
-DIA: 10 ~ 100 mmHg
-MEAN: 20 ~ 110 mmHg
Resolution Pressure: 1mmHg
Accuracy Pressure Maximum Mean error: ±5mmHg Maximum Standard deviation: ±8mmHg Overpressure Protection
-Adult Mode: 297±3 mmHg
-Pediatric Mode: 240±3 mmHg
-Neonatal Mode: 147±3 mmHg
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified Actualization interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate
Measuring and Alarm Range: 20~300bpm Resolution: 1bpm
Accuracy: ±2bpm
Channel: 2
Measuring and Alarm Range: 0 ~ 50 °C Resolution: 0.1°C
Accuracy: ±0.1°C
Actualization interval: about 1 Sec. Average Time Constant:<10 Sec.
Adult finger SpO2 sensor Adult NIBP cuff
NIBP extension tube AC power adaptor Operation instruction
5-lead ECG cable (including RESP) Adult finger SpO2 sensor
Adult NIBP cuff NIBP extension tube TEMP sensor
AC power adaptor ECG electrodes Operation instruction
Central Monitoring System
1. 10.1-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR
2. 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
3. Efficient resistance to interference of defibrillator and electrosurgical cautery;
4. SPO2 can testing for 0.1% Weak;
5. RA-LL impedance Respiration;
6. Trend Coexist Display;
7. OxyCRG DyNamic View Display;
8. Bed to Bed view Display;
9. Networking capacity and nurse calling system;
10. Options of IBP, EtCO2, Central Monitoring System;
11. Built-in rechargeable battery 4400mAh;
12. 10.1" high resolution color TFT LCD display;
13. Large volume of tabular and graphic trends information storage and easy to recall;
14. Anti-ESU, anti- defibrillator;
15. Capture dynamic waveforms.
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V, Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL) Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Range
Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which great Resolution: 1 bpm
Sensitivity: > 200 (uV P-P) Differential Input Impedance: > 5 MΩ CMRR
-Monitor: > 105 dB
-Operation: > 105 dB
-Diagnosis: > 85 dB
Electrode offset potential: ±300mV Leakage Current: < 10 uA
Baseline Recovery: < 3 S after Defi. ECG Signal Range: ±8 m V (Vp-p) Bandwidth
-Surgery: 1 ~ 15 Hz
-Monitor: 0.5 ~ 35 Hz
-Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available Review: Available
Method: Impedance between R-F (RA-LL) Differential Input Impedance: >2.5 MΩ Measuring Impedance Range: 0.3~5.0Ω Base line Impedance Range: 0 – 2.5 KΩ Bandwidth: 0.3 ~ 2.5 Hz
Resp. Measuring Range:
-Adult: 0 ~ 120 rpm
-Neo/Ped: 0 ~ 150 rpm
Resp. Alarm Range:
-High alarm range limit: 8~120 rpm
-Low alarm range limit: 6~120 rpm Resolution: 1 rpm
Accuracy: ±2 rpm
Apnea Alarm: 10 ~ 40 S
Method: Oscillometric Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min Pulse Rate Range: 40 ~ 240 bpm Alarm Type: SYS, DIA, MEAN Measuring and alarm range
Adult Mode
-SYS: 40 ~ 270 mmHg
-DIA: 10 ~ 215 mmHg
-MEAN: 20 ~ 235 mmHg
Pediatric Mode
-SYS: 40 ~ 200 mmHg
-DIA: 10 ~ 150 mmHg
-MEAN: 20 ~ 165 mmHg
Neonatal Mode
-SYS: 40 ~ 135 mmHg
-DIA: 10 ~ 100 mmHg
-MEAN: 20 ~ 110 mmHg
Resolution Pressure: 1mmHg
Accuracy Pressure Maximum Mean error: ±5mmHg Maximum Standard deviation: ±8mmHg Overpressure Protection
-Adult Mode: 297±3 mmHg
-Pediatric Mode: 240±3 mmHg
-Neonatal Mode: 147±3 mmHg
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified Actualization interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate
Measuring and Alarm Range: 20~300bpm Resolution: 1bpm
Accuracy: ±2bpm
Channel: 2
Measuring and Alarm Range: 0 ~ 50 °C Resolution: 0.1°C
Accuracy: ±0.1°C
Actualization interval: about 1 Sec. Average Time Constant:<10 Sec.
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
-ART: 0 ~ 300 mmHg
-PA: -6 ~ 120 mmHg
-CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
-P1/P2: -10 ~ 300 mmHg
Press Sensor
Sensitivity: 5 uV/V/mmHg Impedance: 300-3000Ω Resolution: 1 mmHg
Accuracy: ±2% or ±1mmHg, which great Actualization interval: about 1 Sec
5-lead ECG cable (including RESP) Adult finger SpO2 sensor
Adult NIBP cuff NIBP extension tube TEMP sensor
AC power adaptor ECG electrodes Operation instruction
IBP, EtCO2, Central Monitoring System
1. 12.1-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR
2. 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
3. Efficient resistance to interference of defibrillator and electrosurgical cautery;
4. SPO2 can testing for 0.1% Weak;
5. RA-LL impedance Respiration;
6. Trend Coexist Display;
7. OxyCRG DyNamic View Display;
8. Bed to Bed view Display;
9. Networking capacity and nurse calling system;
10. Options of Internal Recorder, IBP, EtCO2, Central Monitoring System, Touch Screen, VGA, Neonate SPO2
11. Built-in rechargeable battery 4400mAh;
12. 12.1" high resolution color TFT LCD display;
13. Large volume of tabular and graphic trends information storage and easy to recall;
14. Anti-ESU, anti- defibrillator;
15. Capture dynamic waveforms.
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V, Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL) Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Range
-Adult: 15 ~ 300 bpm
-Neo/Ped: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which great Resolution: 1 bpm
Sensitivity: >200 (uV P-P) Differential Input Impedance: > 5 MΩ CMRR
-Monitor: > 105 dB
-Operation: > 105 dB
-Diagnosis: > 85 dB
Electrode offset potential: ±300mV Leakage Current: <10uA
Baseline Recovery: < 3S after Defi. ECG Signal Range: ±8 m V (Vp-p) Bandwidth
-Surgery: 1 ~ 15 Hz
-Monitor: 0.5 ~ 35 Hz
-Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available Review: Available
Method: Impedance between R-F (RA-LL) Differential Input Impedance: >2.5 MΩ Measuring Impedance Range: 0.3 ~ 5.0Ω Base line Impedance Range: 0 ~ 2.5 KΩ
Bandwidth: 0.3 ~ 2.5 Hz Resp. Measuring Range:
-Adult: 0 ~ 120 rpm
-Neo/Ped: 0 ~ 150 rpm Resp. Alarm Range:
-High alarm range limit: 8~120 rpm
-Low alarm range limit: 6~120 rpm Resolution: 1 rpm
Accuracy: ±2 rpm
Apnea Alarm: 10 ~ 40 S
Method: Oscillometric Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min Pulse Rate Range: 40 ~ 240 bpm Alarm Type: SYS, DIA, MEAN Measuring and alarm range
Adult Mode
-SYS: 40 ~ 270 mmHg
-DIA: 10 ~ 215 mmHg
-MEAN: 20 ~ 235 mmHg
Pediatric Mode
-SYS: 40 ~ 200 mmHg
-DIA: 10 ~ 150 mmHg
-MEAN: 20 ~ 165 mmHg
Neonatal Mode
-SYS: 40 ~ 135 mmHg
-DIA: 10 ~ 100 mmHg
-MEAN: 20 ~ 110 mmHg
Resolution Pressure: 1mmHg
Accuracy Pressure Maximum Mean error: ±5mmHg Maximum Standard deviation: ±8mmHg Overpressure Protection
-Adult Mode: 297±3 mmHg
-Pediatric Mode: 240±3 mmHg
-Neonatal Mode: 147±3 mmHg
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified Actualization interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate
Measuring and Alarm Range: 20~300bpm Resolution: 1bpm
Accuracy: ±2bpm
Channel: 2
Measuring and Alarm Range: 0 ~ 50°C Resolution: 0.1°C
Accuracy: ±0.1°C
Actualization interval: about 1 Sec. Average Time Constant:<10 Sec.
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
-ART: 0 ~ 300 mmHg
-PA: -6 ~ 120 mmHg
-CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
-P1/P2: -10 ~ 300 mmHg
Press Sensor Sensitivity: 5uV/V/mmHg Impedance: 300-3000Ω Resolution: 1 mmHg
Accuracy: ±2% or ±1mmHg, which great Actualization interval: about 1 Sec
5-lead ECG cable (including RESP) Adult finger SpO2 sensor
Adult NIBP cuff NIBP extension tube TEMP sensor
AC power adaptor ECG electrodes Operation instruction
Internal Recorder, IBP, EtCO2, Central Monitoring System, Touch Screen, VGA, Neonate SPO2
1. 15.6-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR
2. 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
3. Efficient resistance to interference of defibrillator and electrosurgical cautery;
4. SPO2 can testing for 0.1% Weak;
5. RA-LL impedance Respiration;
6. Trend Coexist Display;
7. OxyCRG DyNamic View Display;
8. Bed to Bed view Display;
9. Networking capacity and nurse calling system;
10. Options of Internal Recorder, IBP, EtCO2, Central Monitoring System, Neonate SPO2
11. Built-in rechargeable battery 4800mAh;
12. 15.6" high resolution color TFT LCD display;
13. Large volume of tabular and graphic trends information storage and easy to recall;
14. Anti-ESU, anti- defibrillator;
15. Capture dynamic waveforms.
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL) Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Range
Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which great Resolution: 1 bpm
Sensitivity: > 200 (uV P-P) Differential Input Impedance: > 5 MΩ CMRR
-Monitor: > 105 dB
-Operation: > 105 dB
-Diagnosis: > 85 dB
Electrode offset potential: ±300mV Leakage Current: < 10 uA
Baseline Recovery: < 3 S after Defi. ECG Signal Range: ±8 m V (Vp-p) Bandwidth
-Surgery: 1 ~ 15 Hz
-Monitor: 0.5 ~ 35 Hz
-Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available Review: Available
Method: Impedance between R-F (RA-LL) Differential Input Impedance: >2.5 MΩ Measuring Impedance Range: 0.3~5.0Ω Base line Impedance Range: 0 – 2.5 KΩ Bandwidth: 0.3 ~ 2.5 Hz
Resp. Measuring Range:
-Adult: 0 ~ 120 rpm
-Neo/Ped: 0 ~ 150 rpm Resp. Alarm Range:
-High alarm range limit: 8~120 rpm
-Low alarm range limit: 6~120 rpm Resolution: 1 rpm
Accuracy: ±2 rpm
Apnea Alarm: 10 ~ 40 S
Method: Oscillometric Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min Pulse Rate Range: 40 ~ 240 bpm Alarm Type: SYS, DIA, MEAN Measuring and alarm range
Adult Mode
-SYS: 40 ~ 270 mmHg
-DIA: 10 ~ 215 mmHg
-MEAN: 20 ~ 235 mmHg
Pediatric Mode
-SYS: 40 ~ 200 mmHg
-DIA: 10 ~ 150 mmHg
-MEAN: 20 ~ 165 mmHg
Neonatal Mode
-SYS: 40 ~ 135 mmHg
-DIA: 10 ~ 100 mmHg
-MEAN: 20 ~ 110 mmHg
Resolution Pressure: 1mmHg
Accuracy Pressure Maximum Mean error: ±5mmHg Maximum Standard deviation: ±8mmHg Overpressure Protection
-Adult Mode: 297±3 mmHg
-Pediatric Mode: 240±3 mmHg
-Neonatal Mode: 147±3 mmHg
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified Actualization interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate
Measuring and Alarm Range: 20~300bpm
Resolution: 1bpm Accuracy: ±2bpm
Channel: 2
Measuring and Alarm Range: 0 ~ 50 °C Resolution: 0.1°C
Accuracy: ±0.1°C
Actualization interval: about 1 Sec. Average Time Constant:<10 Sec.
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
-ART: 0 ~ 300 mmHg
-PA: -6 ~ 120 mmHg
-CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
-P1/P2: -10 ~ 300 mmHg
Press Sensor
Sensitivity: 5 uV/V/mmHg Impedance: 300-3000Ω Resolution: 1 mmHg
Accuracy: ±2% or ±1mmHg, which great Actualization interval: about 1 Sec
5-lead ECG cable (including RESP) Adult finger SpO2 sensor
Adult NIBP cuff NIBP extension tube TEMP sensor
AC power adaptor ECG electrodes Operation instruction
Internal Recorder, IBP, EtCO2, Central Monitoring System, Neonate SPO2